It was reported that during a da vinci-assisted surgical procedure, the "code" on the bipolar energy cable had melted.The procedure was completing as planned with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the customer reported no sparking, smoke, fire, or thermal damage.The cable was inspected prior to the event with no damage noted, and the number of prior uses and how long it had been on site were both unknown.At the time, the cable was connected to an integrated electro surgical unit (iesu) generator using bipolar energy and was plugged into maryland and fenestrated bipolar instruments; no information was provided on whether it was connected to the correct generator port or whether a third-party generator was in use.There were no injuries, the issue was resolved by switching to a backup cable.
|