H11: additional manufacturer narrative:
the dhr review was completed and there are no related ncrs.Engineering evaluation summary:
per (b)(4), stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.There is insufficient information available regarding patient or procedural factors known to cause or contribute to stenosis.Therefore, a definitive root cause cannot be conclusively determined.
|