BOSTON SCIENTIFIC CORPORATION INOGEN? X4 CRT-D; cardiac resynchronization therapy defibrillator (CRT-D) Back to Search Results
Model Number G148
Medical Device Problem Codes Pacing Problem (1439); Defibrillation/Stimulation Problem (1573); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Codes Ventricular Fibrillation (2130); Shock from Patient Lead(s) (3162)
Date of Event 09/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) delivered potential inappropriate therapy for due to atrial undersensing.Technical services (ts) reviewed the stored data and noted the atrial undersensing resulted in the device determined the ventricular rate was greater than the atrial rate, 13 anti-tachycardia pacing (atp) bursts and three electric shocks were delivered, one of the atp bursts accelerated the rhythm to ventricular fibrillation (vf), but the rhythm was ultimately converted.Ts advised evaluation of the atrial lead and recommended reprogramming adjustments.No additional adverse patient effects were reported.This icd remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN? X4 CRT-D
Common Device Name
cardiac resynchronization therapy defibrillator (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50813000
Report Number2124215-2026-49867
Device Sequence Number18819601
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public(01)00802526534591(17)240426
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/26/2024
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number282048
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/18/2026
Date Device Manufactured05/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Life Threatening;
Patient Age62 YR