Manufacturer's investigation conclusion:
the reported event of the centrimag console¿s screen not functioning as intended was not confirmed.The centrimag console (serial number (b)(6) was not returned for analysis.Questions regarding the event were asked multiple times and no responses were received.The root cause of the reported event was unable to be conclusively determined through this analysis.Review of the device history record for the centrimag console, serial number (b)(6), showed the device was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual (rev.M) section 4 "warnings and precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction."
no further information was provided.The manufacturer is closing the file on this event.
|