| Model Number |
U128 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Use of Device Problem (1670); Interrogation Problem (4017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) recorded a code 1003, indicating the voltage is too low for the projected remaining capacity.Additionally, remote monitoring was disabled.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.This device remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|