Model Number M00545900
Medical Device Problem Codes Inflation Problem (1310); Use of Device Problem (1670); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a hurricane rx dilatation balloons dilatation balloon was being prepared to be used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) management procedure for stenosis and pain in lower abdomen performed on (b)(6) 2026.During unpacking, it was noticed that the balloon was damaged and could not be inflated outside the patient.The procedure was completed with another hurricane rx dilatation balloons dilatation balloon.There were no patient complications reported as a result of this event.The patients condition at the conclusion of the procedure was reported to be stable.Note: the instructions for use (ifu) indicate that this balloon should not be pre-inflated prior to use in the procedure.However, the customer reported that the balloon was pre-inflated outside of the patient.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a04 captures the reportable event of balloon damaged/defective.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Hurricane RX
Common Device Name
catheter, biliary, diagnostic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Cork Limited
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key50813127
Report Number3005099803-2026-03848
Device Sequence Number18681350
Product Code FGE
UDI-Device Identifier08714729283812
UDI-Public(01)08714729283812(17)270120(10)35664310
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K130484
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00545900
Device Catalogue NumberM00545900
Device Lot Number0035664310
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2026
Initial Report FDA Received Date09/18/2026
Date Device Manufactured01/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient Weight154 LB
-
-