| Model Number |
CC60WF |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A non-healthcare professional reported that during the intraocular lens (iol) implant procedure lens haptic found to be torn.The lens was implanted and removed on the same day.Additional information received and stated that, the originally scheduled procedure was completed on the same day.As per the surgeon contributory cause was possible issue with the lens cartridge; during injection, the iol trailing haptic stuck on the plunger.When the doctor attempted to free it with his second instrument, the trailing haptic tore off.The tech noted that the cartridge was cracked and indicated that the trailing haptic might have been stuck between the plunger and cartridge.There was no impact to the patient as the doctor was able to remove the damaged iol immediately without issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.3., h.6.And h.11.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|