| Model Number |
470056-11 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to the start of a da vinci-assisted surgical procedure, the 0-deg endoscope plus had a broken tip.The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The endoscope plus has been returned and evaluated by the failure analysis (fa) team.The endoscope was visually inspected and was found with damage at the distal end.The distal window located at distal tip was visually inspected and found cracked.Additionally, the distal window located at the distal tip was visually inspected and found to have a broken or detached piece in a location that could fall into the patient.Also, the endoscope was placed through friction test using a screening aide to manually rotate the adapter to detect for friction.The camera instrument adapter did not rotate properly and/or noises were heard during testing and confirmed as binding.The camera instrument adapter removed from endoscope housing and evaluated and found with aea shaft bearing contributing to friction.The endoscope was visually inspected and confirmed hairline crack was found on the bezel of the button pack assembly.During inspection of the endoscope cable integrated connector, the cable integrated connector housing exhibited discoloration.The complaint was confirmed by failure analysis.
|
| |
|