| Model Number |
470183-19 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the staff noticed smoke coming from the plastic housing unit of the permanent cautery hook instrument.Staff clearly noted smoke was coming from the housing unit and not the tip of the instrument.They removed the instrument and they said you could smell burnt plastic.The procedure was completed with no reported injury.Isi followed up with the initial reporter and obtained the following additional information: no arcing observed during the last procedure in which the instrument was used.There was no injury to the patient.Smoke was not coming from the tip.The product has already been shipped back to isi for evaluation.No photographic images of the instrument available for review.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The permanent cautery hook instrument was analyzed and found to have thermal damage on the monopolar yaw pulley at the distal end.Material appears to have burned off from the charring that occurred near the yaw hook/spatula.The instrument passed the electrical continuity test.The instrument passed the energy delivery test on 3 out of 3 attempts.During the energy delivery test, smoke came from the yaw pulley due to the thermal damage.Visual inspection found no damage to the conductor wire or its insulation.
|
| |
|