Datascope Corp. Sensation Plus 7.5Fr. 40cc IAB with Accessories, US Only; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0684-00-0568-01U |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by direct call from the customer to the emergency support program, that during use of sensation plus 40cc balloon catheter the inner lumen of the iab catheter was clotted; customer was advised to remove the fluid-filled pressurized bag from the inner lumen and label it "do not use" due to the risk of thrombus.The fiber-optic (fo) cable was being used as the arterial pressure source.The customer reported a lead fault on a cardiosave iabp in auto mode, intermittently switching to pressure trigger.The customer was advised to replace the ecg electrodes, apply doppler gel, and reposition the electrodes closer to the heart to improve conduction.At the time of the call, the ecg waveform was clean, and the customer agreed to follow the recommendations if needed.No harm to the patient reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was conducted regarding clotted inner lumen in the sensation plus 7.5fr.Balloon clotted inner lumen is a blockage of the catheter¿s inner channel, and it is typically caused by blood clot (thrombus) formation within that lumen.When the inner lumen becomes occluded by a blood clot, it can no longer transmit accurate arterial pressure.The clotted inner lumen can occur due to one or more of the following factors:
¿ inadequate flushing of the inner lumen.¿ kink in the inner lumen.¿ iab remaining inactive/standby for more than 30 minutes.These conditions may promote a slow blood flow within the inner lumen, leading to thrombus formation and occlusion.Once clotted, the inner lumen can no longer transmit accurate arterial pressure waveforms.Per ifu guidelines, failure to aspirate and flush the lumen immediately after catheter placement or when waveform damping is observed increases the risk of clot formation.In cases where resistance is met during aspiration or flushing is unsuccessful, the lumen should be considered occluded and discontinued from use.This issue continues to be monitored within the manufacturer¿s quality system.
|
| |
|