| Model Number |
0998-00-0800-45 |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the fse that during a preventive maintenance the cardiosave intra-aortic balloon pump (iabp) had a non-functional power switch.The device powers on after being plugged in.The machine was displaying an error code 139.There was no patient involved and no harm is reported.A getinge fse was dispatched to evaluate the unit.The fse reported that after power management board was replaced the error was resolved.Moreover, it was reported that the problem remained and due to the shortage of spare parts it has not been resolved and the equipment was reported to be unusable.The equipment has been delivered to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site name: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|