• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Diabetes Care Inc FreeStyle Libre 3 Plus; Continuous Glucose Monitoring System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Diabetes Care Inc FreeStyle Libre 3 Plus; Continuous Glucose Monitoring System Back to Search Results
Model Number 78768-01
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported receiving erroneous glucose results from an abbott diabetes care device.The customer received sensor scan results of 59 mg/dl compared to glucose values of 283 mg/dl respectively obtained on the comparison device, when plotted on a parkes error grid, fall into the c zone, showing the difference in values to be clinically significant.The length of sensor wear was 1 day.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
The reported device was returned and investigated.The investigation included a review of the complaint information, available device data, and an inspection of the returned product.A visual inspection of the device was performed to assess for evidence of physical abnormalities.Functionality testing was performed to evaluate device performance to ensure device is functioning as intended.The device history records (dhrs) for the product were reviewed and the dhrs showed the product passed all tests prior to release.Based on the investigation findings, no device malfunction or product deficiency was identified; therefore, the issue is not confirmed.This complaint was reviewed and is being submitted as it has been reassessed as reportable as the result of retrospective review associated with a capa.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FreeStyle Libre 3 Plus
Common Device Name
Continuous Glucose Monitoring System
Manufacturer (Section D)
Abbott Diabetes Care Inc
1360 south loop road
alameda CA 94502 -700
Manufacturer Contact
melissa pieplow
1360 south loop road
alameda, CA 94502--700
+1(510)384-5975
MDR Report Key50814241
Report Number2954323-2026-421919
Device Sequence Number10127435
Product Code QBJ
UDI-Device Identifier30357599843008
UDI-Public(01)30357599843008(17)2025-01-31(10)T60002214(91)78769-01
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K223435
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2025
Device Model Number78768-01
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer11/01/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2024
Initial Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-