A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 09-apr-2026 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the pinched drawer sensor cable and latch inseparable was faulty.A field service engineer (fse) performed troubleshooting on the auxiliary cabinet full height drawer that was displaying a drawer not sensing closed message.Inspection identified a pinched drawer sensor cable.The drawer sensor was replaced, and the inseparable latch was also replaced.Hardware test application (hta) testing confirmed the drawer was functioning correctly, but the application continued to indicate the drawer as failed.The rear chassis was inspected, no adverse conditions were identified, and all status lights were operating normally.The system functioned as intended after the field service engineer repaired the device.
|