Model Number VH-4000
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that when harvester had completed dissection and was taking branches he did experience the vasoview hemopro 2 shutting down after some prolonged burning, but the it did not recover at all with rest.The power supply was changed out with no effect.At that point, a new hemopro 2 was opened and that one worked as expected.Power supplies set at 3.There was an audible beep from the power supply during activation when the difficulty cutting branches occurred.They were unable to determine whether both lights on the power supply were illuminated.The toggle was fully slid backward to deliver energy to the jaws.The hp2 does not have a tactile click to indicate activation.There was a delay due to switching out the components and finally having to open another kit.There was no patient harm or adverse effects.
 
Additional Manufacturer Narrative
Tw (b)(4) event site name exceeds character limitations: (b)(6).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VasoView Hemopro 2
Common Device Name
Electrosurgical, Cutting & Coagulation & Accessories
Manufacturer (Section D)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key50815001
Report Number2242352-2026-0000812
Device Sequence Number10218574
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public(01)00607567700406(10)3000574938(17)280706
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K101274
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Device Lot Number3000574938
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/11/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/07/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
No concomitant medical products needed.
Patient Age67 YR
Patient SexMale
Patient Weight156 LB
-
-