A philips remote technical support engineer (tse) interviewed the customer and recommended nbp calibration, accuracy verification, leak testing, and advised replacement of the nbp pump assembly if testing failed or the issue persisted.The customer¿s biomed representative agreed to perform the recommended testing and obtain a replacement nbp pump assembly if necessary.Part number information was provided.Based on the remote assessment, the nbp pump assembly may require calibration or replacement.The customer assumed responsibility for further servicing of the device.No additional information is expected, as philips remote support for this event has concluded.The product malfunction was unable to be confirmed because the customer resolved the issue on their own.A review of the service history for this device shows the problem has not been reported again for this device.If additional information is received, the complaint file will be reopened for further investigation.
|