Datascope Corp. Sensation Plus 7.5Fr. 40cc IAB with Accessories, US Only; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0684-00-0568-01U |
| Medical Device Problem Codes |
Inappropriate Waveform (2536); Device Sensing Problem (2917); No Pressure (2994)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Updated fields: b4, e1, g3, g6, h2, h11\
correction fields: b1, d10, g7, h6(type of investigation, investigation findings, investigation conclusions), please revert a1 and d8 to blank as these details were sent inadvertently in initial report.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site name: (b)(6).\
\
difficult/unable to monitor arterial pressure is a condition in which the intra-aortic balloon pump (iabp) cannot obtain a reliable arterial pressure waveform due to issues such as sensor failure (fiber optic iab calibration failure, disconnection, or fiber optic cable break), catheter lumen obstruction (e.G., clot, kink, or lumen damage), setup or equipment problems (e.G., improper zeroing, air bubbles, damping, or loose connections), or patient-related factors such as low pulse pressure or arrhythmias.In these situations, the arterial pressure signal may become inaccurate or unavailable for safe iabp operation.\
difficult/unable to monitor arterial pressure can occur due to one or more of the following factors:\
- blood clot (thrombus) formation within the inner lumen,\
- break in the inner lumen,\
- kink in the inner lumen,\
- fiber optic iab calibration failure or disconnection,\
- fiber optic cable break,\
- fiber optic cable kink,\
- transducer not properly zeroed or vented,\
- overdamped or flat arterial pressure waveform due to air bubbles or excessive tubing compliance,\
- loose or incorrect pressure cable connections,\
- off-label use.\
\
a sensor failure can occur due to one or more of the following probable causes:\
- optical sensor kink,\
- break in the sensor,\
- connector issue.\
\
this issue continues to be monitored within the manufacturer's quality system.
|
| |
|
Event or Problem Description
|
|
It was reported by direct call from the customer to the emergency support program, that during use of sensation plus 40cc balloon catheter a poor-quality arterial waveform on the iabp monitor while using the fiberoptix (fo) signal as the ap source.The fo arterial waveform displayed a whip-like artifact with pressure readings of 52/20 mmhg (map 44) and augmentation of 62.Comparison with the bedside monitor using the radial arterial line showed stable pressures of 104/61 mmhg (map 80).Troubleshooting, including multiple flushes of the inner lumen, did not resolve the waveform artifact.Review of a chest x-ray confirmed the distal radiopaque marker was positioned between the 2nd and 3rd intercostal spaces.Technical support recommended switching the ap source from fo to the transducer and guided the clinician through the process.Following the change, appropriate arterial waveforms and pressure values were obtained.No patient harm or injury was reported.
|
| |
|