It was reported by the customer that during routine check the cardiosave intra-aortic balloon pump (iabp) shut down.No patient involved.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that a review of the system logs found no anomalies, alerts, or prior error codes that would account for the reported shutdown.The device underwent inspection and functional testing, including battery charging/discharging verification, battery operation checks, error review, and confirmation of the most recent battery maintenance ((b)(6)2026).All tests were successfully completed, the device remained operational during testing with no shutdowns observed, and no issues were identified.The equipment is functioning correctly and is suitable for clinical use.
|