The account reported finding adhesive issue with sure site.A sample was received for the investigation; the account returned 5 dressings, all the dressings provided had been peeled, and the liner was partially or completely removed from the dressings with the film stuck to the backing.All samples confirm the issue reported by the account.The root cause is considered to be a vendor product issue relating to the component manufacturing process; the issue derives from the raw materials process when the force between frame liner and the film is too weak, and the complaint has been provided to the component manufacturing facility for further evaluation.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|