Gyrus ACMI, Inc. EZDILATE BALLOON DILATOR (WG) 18-19-20 Back to Search Results
Model Number BD-410X-2055
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the balloon broke while inside of the patient involving an olympus dilator.The issue occurred during esophagogastroduodenoscopy with dilation procedure.No information regarding whether any pieces broke off inside the patient or if retrieval was required.The procedure was completed with an olympus back-up device.There was no patient reported harm.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection, and the reported failure was confirmed.Based on the results of the investigation, the probable cause of the reported failure is excessive mechanical stress applied to the device during use.A device history record review was completed, and no issues were identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZDILATE BALLOON DILATOR (WG) 18-19-20
Common Device Name
EZDILATE BALLOON DILATOR (WG) 18-19-20
Manufacturer (Section D)
Gyrus ACMI, Inc.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
Gyrus ACMI, Inc.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50816902
Report Number3011050570-2026-00917
Device Sequence Number13301398
Product Code FGE
UDI-Device Identifier00821925033276
UDI-Public00821925033276
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K180086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBD-410X-2055
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/03/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/21/2026
Initial Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-