| Model Number |
CM1100 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Thermal damage of the pcb board was discovered during evaluation of the device.The patient electronic unit was replaced and there were no adverse consequences reported by the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One cardiomems patient electronic system was received for evaluation.Visual inspection verified thermal damage on the pcb board.The damage to pcb cascaded in the unit not powering on at all.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified that are related to the reported event.
|
| |
|