Model Number 9-ACP2-010-031
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026, a 31mm amplatzer amulet left atrial appendage occluder was implanted in the patient.On follow up transthoracic echocardiogram (tte) prior to discharge, it was noted that the amulet device had embolized to the left ventricle (lv).The patient was stable without symptoms.The patient underwent a procedure to retrieve the embolized device via a retrograde aortic approach.The device was retrieved successfully.The patient remained hemodynamically stable throughout the time the device was discovered to be in the lv and was retrieved.There was no damage to the mitral valve or aortic valve was noted.There was no change to the patient's heart function post device retrieval was noted.The patient was kept overnight for monitoring.
 
Additional Manufacturer Narrative
A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Amplatzer Amulet
Common Device Name
Cardiac Plug
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
St. Jude Medical #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50817237
Report Number2135147-2026-06291
Device Sequence Number21507070
Product Code NGV
UDI-Device Identifier00811806013510
UDI-Public(01)00811806013510(17)270930(10)8722960
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P200049
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number9-ACP2-010-031
Device Catalogue Number9-ACP2-010-031
Device Lot Number8722960
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexMale
-
-