| Model Number |
M0061752630 |
| Medical Device Problem Codes |
Break (1069); Deformation Due to Compressive Stress (2889); Detachment of Device or Device Component (2907); Material Deformation (2976)
|
| Health Effect - Clinical Codes |
Abdominal Pain (1685); Foreign Body In Patient (2687)
|
| Date of Event |
05/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a percuflex plus ureteral stent was implanted into the right ureter of the patient during a double j (jj) stent placement performed on, (b)(6) 2026.Post procedure, on the same date, the patient experienced abdominal pain.The following day, (b)(6) 2026, x-ray imaging revealed that the stent had broken inside the patient.On (b)(6) 2026, endoscopic surgery was performed to remove the broken stent, however, the coil could not be removed and was left unretrieved.The patient's condition was stable at the conclusion of this procedure, however, further hospital observation was required.On (b)(6) 2026, another endoscopic surgery was performed and successfully retrieved all broken fragments.There was no information if a new stent was replaced and there was no information if medication was given for pain, however, it was reported that the adverse event of pain was resolved.The patient's condition following the procedure was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Block a2:
patient's exact age is unknown; however, it was reported that the patient's year of birth was 2001.Block d4, h4:
the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete udi and other product specific information.If additional details become available, a supplemental report will be submitted.Block h6:
patient code e2008 captures the reportable event of unretrieved device fragment.Impact code of f23 captures the reportable event of unexpected medical intervention.Impact code of f08 captures the reportable event of prolonged hospitalization.Device code of a0501 captures the reportable event of coil detachment inside the patient.Device code a040601 captures the reportable event of stent buckled material inside the patient.Device code a0406 captures the reportable event of stent material deformation inside the patient.Device code a0401 captures the reportable event of stent shaft break.Block h11: investigation analysis.The device was not returned for analysis; therefore, a technical analysis could not be performed.However, a photo was of the percuflex plus ureteral stent was provided and showed that the stent shaft was detached in six pieces, the renal coil was detached, and parts of the stent were buckled and kinked.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A risk review was completed and confirmed that the event of coil detachment of device or device component, stent buckled material, stent material deformation and stent shaft break, and, pain, abdominal and unretrieved device fragments were defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Additionally, the impact code of imaging required, unexpected medical intervention, and hospitalization or prolonged hospitalization, are subsequent event of the primary anticipated event.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to procedure.
|
| |
|
Search Alerts/Recalls
|
|
|