| Model Number |
i-Res Warmer 1.0 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a malfunction during service causing loss of oxygen concentration.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The check valve assembly was replaced to resolve the issue.Block a: no report of patient involvement.Legal manufacturer:
gehc trout - 3114 n grandview blvd.Usa waukesha, wi 53188.
|
| |
|