| Model Number |
PFA-GEN-01 |
| Medical Device Problem Code |
Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise of thermal damage identified during device evaluation.During a pulsed field ablation atrial fibrillation procedure, a loud bang was heard from the current pfa generator, and a pfa generator hardware error appeared on the screen.It was noted that mapping of atrial tachycardia in the left atrium had begun, and the capacitors were charged and ready for pfa delivery, but no pfa was delivered when the loud bang occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
One current pfa generator (sn: (b)(6) was received for evaluation.Based on the investigation and information provided to abbott, the root cause of this issue was determined to be thermal damage to the pinnacle bfc pca and pinnacle bridge board.The batch record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
|
| |
|