The account reported finding the sample "mask cannot connect to bag," and the complaint noted the issue did not occur during use and no injury occurred.A sample was received for investigation; it was visually inspected and compared to a reference unit, and the mask connector was observed to be deformed and not circular in shape, preventing attachment to the resuscitation bag.Medline could confirm the product failed to meet specifications.Medline could not confirm the cause of the issue, and no trends were identified.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|