| Model Number |
SGC0801 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4 degenerative mitral regurgitation (mr), prolapsed posterior leaflet for a mitraclip procedure.During the preparation, stopcock on the dilator would not stay on the post.Second stopcock was opened and would not stay on the dilator port.Stopcock would twist off on its own despite tightening it onto the port.Device was replaced and continued the procedure.The final mr was grade 1.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Search Alerts/Recalls
|