The reported issue was that the cuff wont go down past 200 on the meter.Medline received one handheld aneroid in used condition, visually inspected the sample, saw no manufacturing issues preventing the sample to work properly, and tested the sample according to the instructions for use.The complaint was not confirmed; the sample was inflated up to 300 mmhg, the cuff deflated with no issues, the dial did not get stuck at 200 mmhg, returned to 0 mmhg like normal, and no issues were experienced.The cause was documented as not confirmed with received sample, and the sample worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|