The account reported finding defective component, and the issue occurred during use with no injury reported.A photo was received for investigation; the customer provided a photo for review, finished good production records were reviewed, the bill of materials (bom) was reviewed, and component trending was reviewed.Based on the complaint details, photo evaluation, and investigation, the complaint is confirmed; review of the photo confirmed that the needle hub on one of the needles appeared to be warped or crushed.Based on the available information and the lack of additional details, the point at which the defect occurred could not be determined, and a specific root cause could not be identified.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|