It was reported that the sealing of the custom hand packs and photos of slipped/misplaced/separated thl's; the account reported finding tear here label slipped/misplaced.A photo was received for the investigation, and in the photos it was observed that both tear here labels were no longer securing the end folds of the back table cover; finished good production records and finished good pack build were reviewed, and no other issues were identified with the pack.Based on the photos provided, medline confirmed that the tear here labels had slipped, and the cause was documented as supplier error/mfg: material composition non-asia (medline brnd).Tear here labels do not establish sterility and are intended to hold the pack contents together, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|