| Catalog Number |
DYNJAAHME9 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "stanford health has identified a cracked (b)(6) in their custom anesthesia circuit kit, item (b)(6)" and "they found the fhme cracked inside the packaging.".
|
| |
|
Additional Manufacturer Narrative
|
|
The customer stated that they found the fhme cracked inside the packaging.Medline quality received one unit and visually inspected the sample with no original packaging.The damaged of the cracked filter was confirmed.Based on the complaint details and sample, the issue most likely occurred due to an outside source at some point during production, transit, usage, or storage; the quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|
| |
|