Based on x-ray photographic evidence of a fractured ar-1588tb-1 button, the complaint allegation is confirmed.The radiographic image shows the implanted button separated into two fragments, consistent with a postoperative device fracture.Based on the information provided ¿ which may include the returned device (if available), photographs, videos, event description, and any additional field input ¿ arthrex has determined the most likely cause.Given that the reported event occurred postoperatively following implantation, the most likely contributing factor is patient-specific postoperative loading conditions.Per dfu-0147-eo revision 1, e.Warnings:
9.Postoperatively and until healing is complete, fixation provided by this device should be considered as temporary and may not withstand weight bearing or other unsupported stress.The fixation provided by this device should be protected.The postoperative regimen
prescribed by the physician should be strictly followed to avoid adverse stresses applied to the device and bone.The device dfu also identifies implant bending, loosening, and/or breakage as a potential complication associated with implanted fixation devices.
|