During a volt pvi procedure, a pericardial tamponade requiring sternotomy resulting in death was noted.Transseptal access was obtained using an agilis sheath and brk needle.The lspv was accessed with a non-abbott (merit) inqwire guidewire, with the volt catheter being advanced to the ostium of the lspv.The volt balloon was inflated with 10 cc of saline and contrast medium, and respiratory compensation for catheter movement was performed.The baseline calibration was conducted by rotating the balloon and moving it forward and backward.The first pfa application was initiated, reach 90% completion and the system automatically aborted the application with the error message stating "pfa generator fault - therapy pathway resistance exceeded the acceptable limit".The catheter was disconnected and reconnected, confirmed to be fully inflated and confirmed the guidewire was positioned within the pulmonary vein and not in contact with the splines.Baseline was repeated and the second application was delivered 100% with auto marks.Two subsequent applications were automatically aborted by the system.The generator was restarted, the pfa unified cabled and pfa gen connect cables were disconnected and reconnected.Baseline was performed again and applications 5 and 6 in the lspv were delivered successfully.The physician attempted to advance into the lipv approximately 2 minutes after application 6 and the patient developed emerging bradycardia which was paced via the cs catheter.The patient hemodynamic status continued to decline, and a pericardial tamponade was identified with fluoroscopy.A pericardial effusion was attempted and unsuccessful requiring a cardiac surgeon to perform a sternotomy to control the bleeding.The bleeding could not be controlled and the patient died on the procedure table.
|