It was reported that, following an incident unrelated to the implanted intellis pain neurostimulator in which the patient almost fell while exiting the shower but caught themself, the patient no longer felt stimulation and experienced return of pain and loss of pain relief.A low impedance or short issue was reported.Connection check showed red x indicators on electrodes 4, 5, 6, 7, 9, 12, and 13 and values greater than 40,000 with a possible open short involving lead electrodes 0 through 7 during a full in-pain check.The neurostimulator was reprogrammed.The issue was reported as resolved.The patient was alive with no injury.The manufacturer representative (rep) indicated they would send any further information as they become aware.
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|