The customer reported receiving a medline blood culture luer lock access device that was leaking during use; the customer also reported there was no impact to the patient or user and no serious injury, medical intervention, or follow up care related to the reported incident.Medline received 1ea bcllad blood culture luer lock access device in used condition; visual inspection noted that the sheath was fully retracted, covering the needle, but the sample had been used and was covered in blood, so no functional test could be performed, and reserve stock was ordered instead.When reserve stock was received, 29 samples were visually and functionally inspected, with no indications of manufacturing defects and no indications of leakage during or after testing.The complaint was not confirmed, no root cause was identified, and the samples worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|