It was reported that the grater device was too dull, the handles were too stiff, and the coating on the trays was peeling.The event did not occur during surgery.It was reported that during in-house engineering evaluation, it was determined that the device was deformed/bent.All available information has been disclosed.
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H3, h6:
a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code a0202 provided is not a valid finished goods lot number.The product was returned to depuy synthes for evaluation.Visual analysis of the returned quickset ace grater handle found that the hudson adaptor presents signs of heavy wear and slightly deformations.Additionally, the sleeve of the handle also exhibits deterioration.The assessment was performed manually and using the returned mating device and was not able to replicate the reported condition, the locking mechanism functioned as intended without difficulties.The observed deformed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was not confirmed as the observed condition of the quickset ace grater handle would have not contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|