It was reported that on (b)(6) 2026, a 29mm navitor vision valve was selected for implant using a large flexnav delivery system (ds).The patient did not have a history of right bundle branch block (rbbb), left bundle branch block (lbbb) and 1st/2nd atrioventricular block (av) block.The length of the membranous septum was 2.8mm.There was no calcification extending beneath the aortic annular plane in the interventricular septum.Upon initiation of sedation, the patient was developed with bradycardia.Pre-implant balloon aortic valvuloplasty (pre-bav) was performed by using an unspecified balloon subsequently developed with a complete atrioventricular block (cavb).During the procedure, the valve was able to be implanted at a depth of 3mm on the non-coronary cusp (ncc) and 6mm on the left-coronary cusp (lcc) but the cavb remained unchanged, and a temporary pacemaker was left in place.The patient remained hemodynamically stable throughout the procedure and there was no clinically significant delay reported.The implanter classified this event as being related to the patient's past medical history.The patient was in a stable condition and subsequently discharged.
|