The customer reported several sers complaints about this tubing being cracked where the syringe connects.Medline received 2 cs small bore iv extension set, lot numbers were provided, several samples were provided from the customer, a stock check has been performed, and investigation included review of recent complaint trends.The complaint was confirmed.This issue arises from an incompatibility with the material used to manufacture the device and high content alcohol cleaning solutions; this issue has since been addressed, newer lots have been updated, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|