| Model Number |
5076-52 |
| Medical Device Problem Codes |
Pacing Problem (1439); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient presented with chest pain, and upon review it was found that the right ventricular (rv) lead triggered a lead integrity alert (lia) due to increase in sensing integrity counter (sic) and two or more high rate-non sustained episodes.An rv lead fracture was suspected.The lead exhibited high undefined superior vena cava (svc) coil impedance, varying pacing impedance, oversensing resulting in inhibition of pacing, noise, and high capture thresholds.All high voltage therapies were programmed off.It was also reported that the right atrial (ra) lead exhibited undersensing during atrial tachycardia/atrial fibrillation (at/af) stored episodes resulting in termination of at/af detected event(s) in progress, with inappropriate atrial pacing note at times.The leads remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: ddbb1d1 icd, implanted: (b)(6) 2016 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|