Alcon Grieshaber AG GRIESHABER REVOLUTION DSP Forceps; Forceps, Ophthalmic Back to Search Results
Catalog Number 705.43P
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported that during a right eye vitrectomy, the tips of the ophthalmic forceps remained in the locked position even when the surgeon was not depressing the handle during an intraocular lens exchange procedure.The surgeon opened a second forceps but it also became stuck in the locked position.Two forceps received from same lot number.As a result, a third forceps was opened and successfully used to complete the procedure.The surgery was completed using a total of three end grasping forceps, and no patient harm was reported.This report is pertaining to first of two reports received from the initial reporter.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER REVOLUTION DSP Forceps
Common Device Name
Forceps, Ophthalmic
Manufacturer (Section D)
Alcon Grieshaber AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
Alcon Grieshaber AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
+1(800)757-9780
MDR Report Key50819592
Report Number3003398873-2026-00199
Device Sequence Number18692461
Product Code HNR
UDI-Device Identifier07612717070677
UDI-Public(01)07612717070677(17)290228(11)260326(10)180YW2
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number705.43P
Device Lot Number180YW2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/18/2026
Date Device Manufactured03/26/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexMale
-
-