Additional information regarding the serial number of the device used was requested but was not provided.Analysis was performed by onsite service personnel which included functional testing.The results of the analysis revealed main speaker was misconfigured in windows.The customer was using an image and sound transmission system via hdmi to lan converters to repeat the signal which was leading to a misconfiguration in windows.Based on the results of the analysis, the product was confirmed to be operating per specifications, and the cause of the reported problem was misconfiguration related to the sound and image relaying system used.The issue was resolved by reassigning the main speaker and restarting the system.The product was tested, verified to be operational and is back in service.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.
|