According to the reporter, during a laparoscopic distal pancreatectomy, the handle/trigger was damaged prior to patient use.Although the handle was squeezed, the cartridge jaws were unable to close on the tissue.The jaws could not be closed, and the handle returned to its original position.It was noted that the issue occurred on three handles.A new handle was used to resolve the issue.There was no patient injury.
|
|
D10 concomitant products: egiaushort, egiaushort endogia ultra univ sht stap, (lot # p5l1198); sigtrsb60axt, sigtrsb60axt 60mm xtr thk reinforced rel, (lot # n5h1796y); egiaushort, egiaushort endogia ultra univ sht stap, (lot # p5l1198) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|