Model Number 4000
Medical Device Problem Code Pumping Problem (3016)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the pump displayed a travel coarse error, which is a high-priority alarm that could interrupt therapy if it recurs during use.The issue occurred during an occlusion test in which a stopcock was attached to the end of the syringe and the pump was allowed to run until the error was triggered.There was no patient involvement, and no patient harm was reported.
 
Additional Manufacturer Narrative
No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filled as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Medfusion 4000 Syringe Infusion Pump
Common Device Name
Pump, Infusion
Manufacturer (Section D)
ICU Medical, Inc.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
+1(224)706-2300
MDR Report Key50819690
Report Number3012307300-2026-08799
Device Sequence Number18819646
Product Code FRN
UDI-Device Identifier10610586040818
UDI-Public(01)10610586040818(11)141114
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K111386
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number4000
Device Catalogue Number4000-0101-51
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/14/2014
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-