It was reported that on (b)(6) 2026, an 18mm amplatzer septal occluder was chosen for implant using an unknown delivery system to treat atrial septal defect (asd).During the procedure, upon release of device, the physician noted that the occluder shifted to the left atrial side of the aortic rim while remaining stable on the remaining rims.When the physician attempted to retrieve the device using a snare, the occluder was advanced completely into the left atrium.The device subsequently migrated into the left ventricle and then into the aorta.Femoral arterial access was obtained, and the physician successfully snared and retrieved the device.The patient was reported to be stable throughout the event.
|