| Model Number |
WB91051W |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that when using the universal output of the electrosurgical generator, the message "no instrument connected" appeared on the screen during an unspecified procedure and therefore, switched to a different device.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection and the customer allegation was confirmed.The customer allegation is related to error message e486 and was caused by a defective motherboard.The high frequency leakage current, failed current and power consumption and failed power output are consequences of the defect motherboard as root cause.Based on the results of the investigation, the most probable cause of the reported issue is traced to the device.However, a definitive root of reported issue could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|