| Model Number |
D533 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) system exhibited high out of range pacing impedance measurements greater than 3000 ohms in the right atrial (ra) channel.It was also observed that there was a signal artifact monitor (sam) event due to oversensed non-physiologic noise.Atrial tachy response (atr) events were also recorded.Technical services (ts) was contacted and recommended evaluating in office.This icd remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|