Information was received from a patient via a manufacturing representative (rep) who was receiving fentanyl (2000 mcg/ml at 96.2 mcg /day) via an implantable pump for malignant pain and head/neck (non-malignant).It was reported that the patient came in for a while system pump replacement.In pre-op, the patient informed the rep that they felt a difference in their pain when they weaned them down on their dosing.They also stated that the doses had not been helping as much as when they first had it implanted.The discrepancy reports of the latest refills ranged from 5-10 ml more in the pump than had been expected.The healthcare provider (hcp) was able to aspirate easily off the pump, however upon x-ray with dye intra-op, they saw a kink in the catheter.They decided to replace the catheter and the pump.The issue was resolved at the time of the report.
|
|
Product analysis #(b)(4):analysis information -- 2026-09-15 09:15:24 cst pli# 30 product id# 8637-40 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection, alarm output, motor function, and dispense testing.The returned device passed all testing in the laboratory and no anomalies were identified.Continuation of d10: product id 8784 lot# serial# (b)(6) implanted: (b)(6) 2021 explanted: (b)(6) 2026 product type catheter product id 8782 lot# serial# (b)(6) implanted: (b)(6) 2021 explanted: (b)(6) 2026 product type catheter section d information references the main component of the system.Other relevant device(s) are: product id: 8784, serial/lot #: (b)(6), ubd: 18-mar-2023, udi#: (b)(4) ; product id: 8782, serial/lot #: (b)(6), ubd: 08-jul-2023, udi#: (b)(4) h3: analysis of the catheter "[s/n:(b)(6)]" found no anomalies were noted on microscopic inspection, the catheter was patent, and passed pressure leak testing.H3: analysis of the catheter "[s/n:(b)(6)]" identified a kink in the catheter body.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|