| Model Number |
TCDS0801-XTW |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026 a patient presented with grade 4 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, first triclip xtw was inserted into the right atrium through guide, it appeared that there might be clot on the clip.Clip was removed from the guide for the inspection.Triclip was replaced with the new clip and continued the procedure.All steps were performed as per ifu.The final tr was grade 1.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the information reviewed, a cause for the reported thrombosis cannot be determined.The reported patient effect of thrombsis, as listed in the instructions for use, is a known possible complication associated with mitraclip procedures.The reported unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
|
| |
|