• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical TriClip G5 Delivery System; tricuspid valve repair device, percutaneously delivered

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical TriClip G5 Delivery System; tricuspid valve repair device, percutaneously delivered Back to Search Results
Model Number TCDS0801-XTW
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 08/24/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026 a patient presented with grade 4 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, first triclip xtw was inserted into the right atrium through guide, it appeared that there might be clot on the clip.Clip was removed from the guide for the inspection.Triclip was replaced with the new clip and continued the procedure.All steps were performed as per ifu.The final tr was grade 1.There were no adverse patient effects or clinically significant delays reported.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the information reviewed, a cause for the reported thrombosis cannot be determined.The reported patient effect of thrombsis, as listed in the instructions for use, is a known possible complication associated with mitraclip procedures.The reported unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TriClip G5 Delivery System
Common Device Name
tricuspid valve repair device, percutaneously delivered
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50820036
Report Number2135147-2026-06329
Device Sequence Number10218237
Product Code NPS
UDI-Device Identifier05415067050731
UDI-Public(01)05415067050731(17)270112(10)50714R1003
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P230007
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTCDS0801-XTW
Device Catalogue NumberTCDS0801-XTW
Device Lot Number50714R1003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Triclip steerable guide catheter.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexMale
Patient RaceWhite
-
-