(b)(4) vista nova, patient site id: (b)(6).It was reported that on (b)(6) 2026, a 23mm navitor valve with radiopaque markers was chosen for implant with a small flexnav delivery system (ds).During procedure, the small flexnav ds could not pass through the right common iliac artery (cia) due to severe calcification.A wire exchange was performed after 7mm plain old balloon angioplasty (poba) with a non-abbott balloon.The small flexnav ds was able to pass, however, the valve could not cross the sino-tubular junction due to severe calcification, as well.A snare was used to assist with delivery and successfully implant the valve.Patient remained hemodynamically stable throughout the procedure.Post-procedure after returning to the ward, patient complained of back pain.Computed tomography demonstrated an acute type b aortic dissection extending from the distal aortic arch to the abdominal aorta.The physician attributed the cause of the aortic dissection related to difficulty of advancing the flexnav ds through patient anatomy.Prolonged hospitalization was reported.Patient was administered medication.
|