| Model Number |
TCDS0801-XT |
| Medical Device Problem Code |
Difficult or Delayed Activation (2577)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026 a patient presented with grade 5 degenerative tricuspid regurgitation (tr) for a triclip procedure.During the procedure, while deploying triclip xt, actuator knob was turned 11 times, and the actuator knob was not released from the clip.Actuator knob was released when physician pulled the actuator knob out and the device was implanted.The final tr was grade 1.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|