• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Inc. STRATAFIX; Suture, absorbable, synthetic, polyglycolic acid

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Inc. STRATAFIX; Suture, absorbable, synthetic, polyglycolic acid Back to Search Results
Catalog Number SXMP1B119
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a skin closure procedure on (b)(6)2026 and barbed suture was used.During the procedure this barbed suture code has been breaking as they close the skin.They have been closing with this suture for a long time and never had issues.No delay or patient harm was reported.They noticed that it has been the last couple of cases that they have used the product on.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: 1.If possible, please confirm the number of procedures/patients affected by this issue: there were at least 3 cases.2.How many devices demonstrated the alleged deficiency on each involved patient event? 5 sutures showed the deficiency.3.When did the alleged deficiency occur (removal from package / during handling prior to use on patient/ during passage through tissue / during tying / post-op)? during passage through tissue.4.Were there any patient consequences? none were reported.Device history review
=
> a manufacturing record evaluation was performed for the finished device batch 10e9qz, sxmp1b119-16 and no non-conformances were identified.Mfg.Date - 03/19/2026, exp.Date - 02/29/2028.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Common Device Name
Suture, absorbable, synthetic, polyglycolic acid
Manufacturer (Section D)
Ethicon Inc.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
Ethicon Inc.-San Lorenzo PR
road 183, km 8.3
san lorenzo PR 00754
PR   00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
+1(303)552-6892
MDR Report Key50820243
Report Number2210968-2026-09431
Device Sequence Number15984492
Product Code GAM
UDI-Device Identifier10705031236745
UDI-Public(01)10705031236745(17)280228(10)10E9QZ
Combination Product (Y/N)Y
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K182873
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXMP1B119
Device Lot Number10E9QZ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/19/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-